TY - JOUR
T1 - Assessing the impact of different penalty factors of the Bayesian reconstruction algorithm Q.Clear on in vivo low count kinetic analysis of [11C]PHNO brain PET-MR studies
AU - Ribeiro, Daniela
AU - Hallett, William
AU - Howes, Oliver
AU - McCutcheon, Robert
AU - Nour, Matthew M.
AU - Tavares, Adriana A.S.
N1 - Funding Information:
Original study funded by the Medical Research Council (MRC): MC-A656-5QD30.
Funding Information:
AAST is funded by the British Heart Foundation (FS/19/34/34354). The authors would like to thank Gaia Rizzo (inviCRO UK) for the excellent support with the MIAKAT software.
Publisher Copyright:
© 2022, The Author(s).
PY - 2022/2/20
Y1 - 2022/2/20
N2 - Introduction: Q.Clear is a Bayesian penalised likelihood (BPL) reconstruction algorithm available on General Electric (GE) Positron Emission Tomography (PET)-Computed Tomography (CT) and PET-Magnetic Resonance (MR) scanners. This algorithm is regulated by a β value which acts as a noise penalisation factor and yields improvements in signal to noise ratio (SNR) in clinical scans, and in contrast recovery and spatial resolution in phantom studies. However, its performance in human brain imaging studies remains to be evaluated in depth. This pilot study aims to investigate the impact of Q.Clear reconstruction methods using different β value versus ordered subset expectation maximization (OSEM) on brain kinetic modelling analysis of low count brain images acquired in the PET-MR. Methods: Six [11C]PHNO PET-MR brain datasets were reconstructed with Q.Clear with β100–1000 (in increments of 100) and OSEM. The binding potential relative to non-displaceable volume (BPND) were obtained for the Substantia Nigra (SN), Striatum (St), Globus Pallidus (GP), Thalamus (Th), Caudate (Cd) and Putamen (Pt), using the MIAKAT™ software. Intraclass correlation coefficients (ICC), repeatability coefficients (RC), coefficients of variation (CV) and bias from Bland–Altman plots were reported. Statistical analysis was conducted using a 2-way ANOVA model with correction for multiple comparisons. Results: When comparing a standard OSEM reconstruction of 6 iterations/16 subsets and 5 mm filter with Q.Clear with different β values under low counts, the bias and RC were lower for Q.Clear with β100 for the SN (RC = 2.17), Th (RC = 0.08) and GP (RC = 0.22) and with β200 for the St (RC = 0.14), Cd (RC = 0.18)and Pt (RC = 0.10). The p-values in the 2-way ANOVA model corroborate these findings. ICC values obtained for Th, St, GP, Pt and Cd demonstrate good reliability (0.87, 0.99, 0.96, 0.99 and 0.96, respectively). For the SN, ICC values demonstrate poor reliability (0.43). Conclusion: BPND results obtained from quantitative low count brain PET studies using [11C]PHNO and reconstructed with Q.Clear with β < 400, which is the value used for clinical [18F]FDG whole-body studies, demonstrate the lowest bias versus the typical iterative reconstruction method OSEM.
AB - Introduction: Q.Clear is a Bayesian penalised likelihood (BPL) reconstruction algorithm available on General Electric (GE) Positron Emission Tomography (PET)-Computed Tomography (CT) and PET-Magnetic Resonance (MR) scanners. This algorithm is regulated by a β value which acts as a noise penalisation factor and yields improvements in signal to noise ratio (SNR) in clinical scans, and in contrast recovery and spatial resolution in phantom studies. However, its performance in human brain imaging studies remains to be evaluated in depth. This pilot study aims to investigate the impact of Q.Clear reconstruction methods using different β value versus ordered subset expectation maximization (OSEM) on brain kinetic modelling analysis of low count brain images acquired in the PET-MR. Methods: Six [11C]PHNO PET-MR brain datasets were reconstructed with Q.Clear with β100–1000 (in increments of 100) and OSEM. The binding potential relative to non-displaceable volume (BPND) were obtained for the Substantia Nigra (SN), Striatum (St), Globus Pallidus (GP), Thalamus (Th), Caudate (Cd) and Putamen (Pt), using the MIAKAT™ software. Intraclass correlation coefficients (ICC), repeatability coefficients (RC), coefficients of variation (CV) and bias from Bland–Altman plots were reported. Statistical analysis was conducted using a 2-way ANOVA model with correction for multiple comparisons. Results: When comparing a standard OSEM reconstruction of 6 iterations/16 subsets and 5 mm filter with Q.Clear with different β values under low counts, the bias and RC were lower for Q.Clear with β100 for the SN (RC = 2.17), Th (RC = 0.08) and GP (RC = 0.22) and with β200 for the St (RC = 0.14), Cd (RC = 0.18)and Pt (RC = 0.10). The p-values in the 2-way ANOVA model corroborate these findings. ICC values obtained for Th, St, GP, Pt and Cd demonstrate good reliability (0.87, 0.99, 0.96, 0.99 and 0.96, respectively). For the SN, ICC values demonstrate poor reliability (0.43). Conclusion: BPND results obtained from quantitative low count brain PET studies using [11C]PHNO and reconstructed with Q.Clear with β < 400, which is the value used for clinical [18F]FDG whole-body studies, demonstrate the lowest bias versus the typical iterative reconstruction method OSEM.
KW - Bayesian
KW - Neuroimaging
KW - PET-MR
KW - Reconstruction
KW - [C]PHNO
UR - http://www.scopus.com/inward/record.url?scp=85125358837&partnerID=8YFLogxK
U2 - 10.1186/s13550-022-00883-1
DO - 10.1186/s13550-022-00883-1
M3 - Article
AN - SCOPUS:85125358837
SN - 2191-219X
VL - 12
JO - EJNMMI Research
JF - EJNMMI Research
IS - 1
M1 - 11
ER -