60 Citations (Scopus)

Abstract

Objectives
This study sought to identify the effect of coronary autoregulation on myocardial perfusion during intra-aortic balloon pump (IABP) therapy.

Background
IABP is the most commonly used circulatory support device, although its efficacy in certain scenarios has been questioned. The impact of alterations in microvascular function on IABP efficacy has not previously been evaluated in humans.

Methods
Thirteen patients with ischemic cardiomyopathy (left ventricular ejection fraction: 34 ± 8%) undergoing percutaneous coronary intervention were recruited. Simultaneous intracoronary pressure and Doppler-flow measurements were undertaken in the target vessel following percutaneous coronary intervention, during unassisted and IABP-assisted conditions. Coronary autoregulation was modulated by the use of intracoronary adenosine, inducing maximal hyperemia. Wave intensity analysis characterized the coronary wave energies associated with balloon counterpulsation.

Results
Two unique diastolic coronary waves were temporally associated with IABP device use; a forward compression wave and a forward expansion wave caused by inflation and deflation, respectively. During basal conditions, IABP therapy increased distal coronary pressure (82.4 ± 16.1 vs. 88.7 ± 17.8 mm Hg, p = 0.03), as well as microvascular resistance (2.32 ± 0.52 vs. 3.27 ± 0.41 mm Hg cm s–1, p = 0.001), with no change in average peak velocity (30.6 ± 12.0 vs. 26.6 ± 11.3 cm s–1, p = 0.59). When autoregulation was disabled, counterpulsation caused an increase in average peak velocity (39.4 ± 10.5 vs. 44.7 ± 17.5 cm s–1, p = 0.002) that was linearly related with IABP–forward compression wave energy (R2 = 0.71, p = 0.001).

Conclusions
Autoregulation ameliorates the effect of IABP on coronary flow. However, during hyperemia, IABP augments myocardial perfusion, principally due to a diastolic forward compression wave caused by balloon inflation, suggesting IABP would be of greatest benefit when microcirculatory reserve is exhausted.
Original languageEnglish
Pages (from-to)631-640
Number of pages10
JournalJacc-Cardiovascular Interventions
Volume7
Issue number6
DOIs
Publication statusPublished - 1 Jun 2014

Fingerprint

Dive into the research topics of 'Coronary and microvascular physiology during intra-aortic balloon counterpulsation'. Together they form a unique fingerprint.

Cite this