Estimation of subject coregistration errors during multimodal preclinical imaging using separate instruments: Origins and avoidance of artifacts

Research output: Contribution to journalArticle

Jean Philippe Dillenseger, Christian Goetz, Amira Sayeh, Chris Healy, Isabelle Duluc, Jean Noël Freund, André Constantinesco, Gaëlle Aubertin-Kirch, Philippe Choquet

Original languageEnglish
Article number035503
JournalJournal of Medical Imaging
Volume4
Issue number3
DOIs
Accepted/In press24 Jul 2017
Published22 Aug 2017

King's Authors

Abstract

We use high-resolution μCT data in multiple experiments to estimate the sources of error during coregistration of images acquired on separate preclinical instruments. In combination with experiments with phantoms, we completed in vivo imaging on mice, aimed at identifying the possible sources of registration errors, caused either by transport of the animal, movement of the animal itself, or methods of coregistration. The same imaging cell was used as a holder for phantoms and animals. For all procedures, rigid coregistration was carried out using a common landmark coregistration system, placed inside the imaging cell. We used the fiducial registration error and the target registration error to analyze the coregistration accuracy. We found that moving an imaging cell between two preclinical devices during a multimodal procedure gives an error of about 200μm at most. Therefore, it could not be considered a source of coregistration errors. Errors linked to spontaneous movements of the animal increased with time, to nearly 1 mm at most, excepted for body parts that were properly restrained. This work highlights the importance of animal intrinsic movements during a multiacquisition procedure and demonstrates a simple method to identify and quantify the sources of error during coregistration.

View graph of relations

© 2018 King's College London | Strand | London WC2R 2LS | England | United Kingdom | Tel +44 (0)20 7836 5454