T1 - Evaluation of four oral fluid devices (DDS®, Drugtest 5000®, Drugwipe 5+® and RapidSTAT®) for on-site monitoring drugged driving in comparison with UHPLC-MS/MS analysis
AU - Strano-Rossi, Sabina
AU - Castrignanò, Erika
AU - Anzillotti, Luca
AU - Serpelloni, Giovanni
AU - Mollica, Roberto
AU - Tagliaro, Franco
AU - Pascali, Jennifer P.
AU - Di Stefano, Delfina
AU - Sgalla, Roberto
AU - Chiarotti, Marcello
N1 - Funding Information:
This article has been produced under the project “TOXTEST” financed by the Italian Department of Antidrug Policy, Presidency of Ministers Council .
PY - 2012/9/10
Y1 - 2012/9/10
N2 - New Italian legislation on driving under the influence of drugs considers oral fluid (OF) as a possible alternative drug testing matrix. On this basis, the present research was carried out to evaluate the applicability of four commercial on-site OF drug screening devices, namely DDS®, Drugtest 5000®, Drugwipe 5+® and RapidSTAT®, in a real operative context.Preliminarily trained police officers tested randomly stopped drivers with two different kits side-by-side during roadside patrols. A central laboratory confirmed on-site kits' results by UHPLC-MS/MS analysis of the saliva specimen remaining after the screening analysis. 1025 drivers were submitted to the OF tests: 11.6% were positive for cocaine and metabolites, 11.1% for THC, 6% for amphetamines and amphetamine-type designer drugs and 2.3% for ketamine.The sensitivities of the kits were 81% (RapidSTAT®), 82% (DDS®), 90% (Drugwipe 5+®) and 97% (Drugtest 5000®) for cocaine and 38% (DDS®), 47% (Drugwipe 5+®), 72% (RapidSTAT®) and 92% (Drugtest 5000®) for THC. Drugtest 5000 was the only kit showing an acceptable sensitivity for on-site application. Only Drugtest 5000® and RapidSTAT® could be evaluated for amphetamines and methamphetamines: Drugtest 5000® showed a sensitivity of 100% in the case of amphetamines and 86% for methamphetamines, while RapidSTAT® 90% and 76% respectively. Nowadays, ketamine is not included in the target analytes of any on-site devices, but it was systematically included in the UHPLC-MS/MS confirmatory analysis. To ensure adequate reliability, MS confirmation of on-site OF screening tests is anyway always necessary, due to the presence of a significant number of false positive results even when using the commercial kit with the best performance.
AB - New Italian legislation on driving under the influence of drugs considers oral fluid (OF) as a possible alternative drug testing matrix. On this basis, the present research was carried out to evaluate the applicability of four commercial on-site OF drug screening devices, namely DDS®, Drugtest 5000®, Drugwipe 5+® and RapidSTAT®, in a real operative context.Preliminarily trained police officers tested randomly stopped drivers with two different kits side-by-side during roadside patrols. A central laboratory confirmed on-site kits' results by UHPLC-MS/MS analysis of the saliva specimen remaining after the screening analysis. 1025 drivers were submitted to the OF tests: 11.6% were positive for cocaine and metabolites, 11.1% for THC, 6% for amphetamines and amphetamine-type designer drugs and 2.3% for ketamine.The sensitivities of the kits were 81% (RapidSTAT®), 82% (DDS®), 90% (Drugwipe 5+®) and 97% (Drugtest 5000®) for cocaine and 38% (DDS®), 47% (Drugwipe 5+®), 72% (RapidSTAT®) and 92% (Drugtest 5000®) for THC. Drugtest 5000 was the only kit showing an acceptable sensitivity for on-site application. Only Drugtest 5000® and RapidSTAT® could be evaluated for amphetamines and methamphetamines: Drugtest 5000® showed a sensitivity of 100% in the case of amphetamines and 86% for methamphetamines, while RapidSTAT® 90% and 76% respectively. Nowadays, ketamine is not included in the target analytes of any on-site devices, but it was systematically included in the UHPLC-MS/MS confirmatory analysis. To ensure adequate reliability, MS confirmation of on-site OF screening tests is anyway always necessary, due to the presence of a significant number of false positive results even when using the commercial kit with the best performance.
KW - DUID
KW - Kits evaluation
KW - On-site testing
KW - Oral fluid
UR - http://www.scopus.com/inward/record.url?scp=84864367864&partnerID=8YFLogxK
U2 - 10.1016/j.forsciint.2012.04.003
DO - 10.1016/j.forsciint.2012.04.003
M3 - Article
C2 - 22554872
AN - SCOPUS:84864367864
SN - 0379-0738
VL - 221
SP - 70
EP - 76
JO - Forensic Science International
JF - Forensic Science International
IS - 1-3
ER -