TY - JOUR
T1 - Optimizing electrical efficacy of leadless cardiac resynchronization therapy and leadless left ventricular septal pacing: insights on left and right ventricular activation from electrocardiographic imaging
AU - Wijesuriya, Nadeev
AU - Strocchi, Marina
AU - Elliott, Mark
AU - Mehta, Vishal
AU - De Vere, Felicity
AU - Howell, Sandra
AU - Mannakkara, Nilanka
AU - Niederer, Steven
AU - Rinaldi, Christopher Aldo
N1 - Publisher Copyright:
© 2024 Heart Rhythm Society
PY - 2024/8
Y1 - 2024/8
N2 - Background: Leadless cardiac resynchronization therapy (CRT) is an emerging heart failure treatment. An implanted electrode delivers lateral or septal endocardial left ventricular (LV) pacing (LVP) upon detection of a right ventricular (RV) pacing stimulus from a coimplanted device, thus generating biventricular pacing (BiVP). Electrical efficacy data regarding this therapy, particularly leadless LV septal pacing (LVSP) for potential conduction system capture, are limited. Objectives: The purpose of this study was to evaluate the acute performance of leadless CRT using electrocardiographic imaging (ECGi) and assess the optimal pacing modality (OPM) of LVSP on the basis of RV and LV activation. Methods: Ten WiSE-CRT recipients underwent an ECGi study testing: RV pacing, BiVP, LVP only, and LVP with an optimized atrioventricular delay (LV-OPT). BiV, LV, and RV activation times (shortest time taken to activate 90% of the ventricles [BIVAT-90], shortest time taken to activate 95% of the LV, and shortest time taken to activate 90% of the RV) plus LV and BiV dyssynchrony index (standard deviation of LV activation times and standard deviation of all activation times) were calculated from reconstructed epicardial electrograms. The individual OPM yielding the greatest improvement from baseline was determined. Results: BiVP generated a 23.7% improvement in BiVAT-90 (P = .002). An improvement of 43.3% was observed at the OPM (P = .0001), primarily through reductions in shortest time taken to activate 90% of the RV. At the OPM, BiVAT-90 improved in patients with lateral (43.3%; P = .0001; n = 5) and septal (42.4%; P = .009; n = 5) LV implants. The OPM varied by individual. LVP and LV-OPT were mostly superior in patients with LVSP, and in those with sinus rhythm and left bundle branch block (n = 4). Conclusion: Leadless CRT significantly improves acute ECGi-derived activation and dyssynchrony metrics. Using an individualized OPM improves efficacy in selected patients. Effective LVSP is feasible, with fusion pacing at LV-OPT mitigating the potential deleterious effects on RV activation.
AB - Background: Leadless cardiac resynchronization therapy (CRT) is an emerging heart failure treatment. An implanted electrode delivers lateral or septal endocardial left ventricular (LV) pacing (LVP) upon detection of a right ventricular (RV) pacing stimulus from a coimplanted device, thus generating biventricular pacing (BiVP). Electrical efficacy data regarding this therapy, particularly leadless LV septal pacing (LVSP) for potential conduction system capture, are limited. Objectives: The purpose of this study was to evaluate the acute performance of leadless CRT using electrocardiographic imaging (ECGi) and assess the optimal pacing modality (OPM) of LVSP on the basis of RV and LV activation. Methods: Ten WiSE-CRT recipients underwent an ECGi study testing: RV pacing, BiVP, LVP only, and LVP with an optimized atrioventricular delay (LV-OPT). BiV, LV, and RV activation times (shortest time taken to activate 90% of the ventricles [BIVAT-90], shortest time taken to activate 95% of the LV, and shortest time taken to activate 90% of the RV) plus LV and BiV dyssynchrony index (standard deviation of LV activation times and standard deviation of all activation times) were calculated from reconstructed epicardial electrograms. The individual OPM yielding the greatest improvement from baseline was determined. Results: BiVP generated a 23.7% improvement in BiVAT-90 (P = .002). An improvement of 43.3% was observed at the OPM (P = .0001), primarily through reductions in shortest time taken to activate 90% of the RV. At the OPM, BiVAT-90 improved in patients with lateral (43.3%; P = .0001; n = 5) and septal (42.4%; P = .009; n = 5) LV implants. The OPM varied by individual. LVP and LV-OPT were mostly superior in patients with LVSP, and in those with sinus rhythm and left bundle branch block (n = 4). Conclusion: Leadless CRT significantly improves acute ECGi-derived activation and dyssynchrony metrics. Using an individualized OPM improves efficacy in selected patients. Effective LVSP is feasible, with fusion pacing at LV-OPT mitigating the potential deleterious effects on RV activation.
UR - http://www.scopus.com/inward/record.url?scp=85199802008&partnerID=8YFLogxK
U2 - 10.1016/j.hroo.2024.07.004
DO - 10.1016/j.hroo.2024.07.004
M3 - Article
VL - 5
SP - 551
EP - 560
JO - Heart Rhythm O2
JF - Heart Rhythm O2
IS - 8
ER -