Abstract
Purpose: To estimate the minimum stimulus energy required to stimulate the corneal nerves, using a thermally cooling stimulus. Methods: The localised temperature change produced in the pre-corneal tear film by the air-pulse stimulus of the Non-Contact Corneal Aesthesiometer (NCCA) was analysed using a thermal imaging camera. Assuming that the cornea behaves as a near perfect black-body, the threshold for energy loss was calculated by combining Stefan-Boltzmann's law and Kirchoff's law for imperfections in black-body radiation, taking into account the likely thermal conductivity of the human cornea. Results:The average normal subject has a threshold for detecting a cooling effect when the local corneal tear film surface temperature drops from a mean of 33.2-32.9 degreesC (i.e. 0.3 degreesC), over a 1 mm diameter circular area (0.785 mm(2)), and a stimulus time duration of 0.9 s. This gives rise to a sensation threshold for perceiving a loss of heat energy of 7.0 ergs, or 7.0 X 10(-7) joules. Conclusion: The cornea possesses stimulus specific receptors, which have an exquisite sensitivity to a rapid, thermally cooling stimulus. (C) 2001 The College of Optometrists. Published by Elsevier Science Ltd. All rights reserved.
| Original language | English |
|---|---|
| Pages (from-to) | 407 - 410 |
| Number of pages | 4 |
| Journal | Ophthalmic and Physiological Optics |
| Volume | 21 |
| Issue number | 5 |
| DOIs | |
| Publication status | Published - 2001 |