Abstract
OBJECTIVES. The objectives of this study were to determine the impact of different volume-targeted levels on the work of breathing and to investigate whether a level that reduced the work of breathing below that experienced during ventilatory support without volume targeting could be determined.
METHODS. The transdiaphragmatic pressure-time product, as an estimate of the work of breathing, was measured for 20 infants (median gestational age: 28 weeks) who were being weaned from respiratory support by using patient-triggered ventilation (either assist-control ventilation or synchronous intermittent mandatory ventilation). The transdiaphragmatic pressure-time product was measured first without volume targeting (baseline) and then at volume-targeted levels of 4, 5, and 6 mL/kg, delivered in random order. After each volume-targeted level, the infants were returned to baseline. Each step was maintained for 20 minutes.
RESULTS. The mean transdiaphragmatic pressure-time product was higher with volume targeting at 4 mL/kg in comparison with baseline, regardless of the patient-triggered mode. The transdiaphragmatic pressure-time product was higher at a volume-targeted level of 4 mL/kg in comparison with 5 mL/kg and at 5 mL/kg in comparison with 6 mL/kg. The mean work of breathing was below that at baseline only at a volume- targeted level of 6 mL/kg.
CONCLUSIONS. Low volume- targeted levels increase the work of breathing during volume-targeted ventilation. Our results suggest that, during weaning, a volume-targeted level of 6 mL/kg, rather than a lower level, could be used to avoid an increase in the work of breathing. Pediatrics 2009; 123: e679-e684
Original language | English |
---|---|
Pages (from-to) | E679 - E684 |
Journal | Pediatrics |
Volume | 123 |
Issue number | 4 |
DOIs | |
Publication status | Published - Apr 2009 |
Keywords
- Ventilator Weaning
- Work of Breathing
- Humans
- Respiratory Distress Syndrome, Newborn
- Infant, Newborn
- Intensive Care, Neonatal
- Infant, Premature
- Energy Metabolism
- Tidal Volume